• 云顶集团3118(中国)

    Autofluorescence Microscopy QC for Multimodal Molecular Imaging

    Nathan Heath Patterson, Jeff Spraggins, Angela R.S. Kruse, Katerina V Djambazova, Melissa Farrow, Lukasz Migas, Raf Van De Plas

    Published: 2023-06-20 DOI: 10.17504/protocols.io.e6nvwjm5dlmk/v1
    AI 解读
    浏览 1279

    Abstract

    Autofluorescence QC is performed prior to collecting on-tissue AF. Fluorescence intensity can be tracked for each fluorescence channel (DAPI, EGFP, and DsRed), allowing the microscope performance and the LED light source to be benchmarked and tracked over time.

    Steps

    Autofluorescence Microscopy QC

    1.

    Fluorescence QC is collected prior to any on-tissue AF collection.

    Use fluorescence calibration slides - StarLight Calibration Slides - Bang Laboratories

    DAPI (excitation-360nm, emission-450nm), Dragon green (excitation-480nm, emission-520nm), and Envy green (excitation-525nm, emission-565nm)

    2.

    Load calibration slides into Zeiss AxioScan.Z1 slide scanner

    3.

    Select the same ~ 25 x 25 mm square region in each calibration slide

    Note
    To reproducibly sample the exact region each time, the square region can be saved as a "region of interest", which can be imported for each slide.

    4.

    Acquire fluorescence microscopy data for each calibration slide.

    Note
    Example set-up and relative ranges: DAPI QC : DAPI (ex. 360 nm, em. 450 nm) Calibration Slide LED Light: 385 nm Light Source Intensity: 90% Exposure time: 20 mseGFP QC : Dragon Green (ex. 480 nm, em. 520 nm) Calibration Slide LED Light: 475 nm Light Source Intensity: 90% Exposure time: 60 msDsRed QC : Envy Green (ex. 525 nm, em. 565 nm) Calibration Slide LED Light: 567 nm Light Source Intensity: 90% Exposure time: 250 ms

    5.

    Data Analysis:

    Once data are collected, background pixels are subtracted, and the average intensity of each channel is calculated and monitored over time.

    推荐阅读

    Nature Protocols
    Protocols IO
    Current Protocols
    扫码咨询