Measuring leaf carbon fractions with the ANKOM2000 Fiber Analyzer
Jocelyne Ayotte, Etienne Laliberté
Abstract
Here we describe the standardized protocol used by the Canadian Airborne Biodiversity Observatory (CABO) to measure carbon fractions in leaf samples, using the ANKOM2000 Fiber Analyzer. Prior to the analysis, leaf samples are oven-dried, and then ground with a cyclone mill (2-mm screen). Then, carbon fractions are measured using a sequential digestion procedure to measure neutral detergent fiber (NDF, acid detergent fiber (ADF), acid detergent lignin (ADL). Finally, the residues are ashed in a muffle furnace to determine inorganic recalcitrant materials. This protocol is based on the manufacturer's protocols.
Before start
Leaf samples need the have been dried for 72 h at 65°C and then ground with a cyclone mill (2-mm screen).
Steps
Sample preparation
Use a solvent resistant marker to label the filter bags to be used in the analysis.
Weigh and record the weight of each empty filter bag to the nearest 0.0001 g (in Fulcrum). Zero the balance.
Equipment
| Value | Label |
|---|---|
| Sartorius Practum | NAME |
| Analytical balance | TYPE |
| Sartorius | BRAND |
| 224-1S | SKU |
Place 0.45–0.50 g of prepared sample in up to 23 of the bags and record the weight to the nearest 0.0001 g (in Fulcrum) of each. Avoid placing the sample in the upper 4 mm of the bag.
Include at least one empty bag in the run to determine the blank bag correction (C1).
Using a heat sealer, completely seal each filter bag closed within 4 mm of the top to encapsulate the sample.
Equipment
| Value | Label |
|---|---|
| Impulse Heat Sealer | NAME |
| Heat Sealer | TYPE |
| Uline | BRAND |
| H-163 | SKU |
Spread the sample uniformly inside the filter bags by gently shaking the bags to eliminate clumping.
Neutral Detergent Fiber
Place up to 3 bags on each of eight Bag Suspender Trays (maximum of 24 bags). Stack the trays on the center post of the Bag Suspender with each level rotated 120 degrees in relation to the tray below it. Place the empty 9th tray on top.
Verify that the hot water supply is on (room B-236).
Verify that the drain hose is securely positioned in an empty yellow waste bin.
Attach the Neutral Detergent (ND) solution hose to the Cubetainer and then to Port A on the instrument.
Add 50 ml of tap water in the alpha-amylase dispenser. Add 8.0mL of alpha-amylase and enough tap water to fill the dispenser (until the line). Attach the Amylase Dispenser Assembly to Port B on the instrument.
Keep the pipette and alpha-amylase bottle next to the instrument for step 16.
Weigh 20gof Sodium sulfite and keep it for step 16.
Open the Vessel Lid and insert the Bag Suspender with bags into the Vessel and place the Bag Suspender weight on top of the empty 9th tray to keep the Bag Suspender submerged.
Follow the instructions on the ANKOM2000 display:
Select NDF . (Wait to close the Vessel Lid.)
The screen will show: Insert samples. Press Enter .
Confirm hot water is on (>70 °C) and press START .
After the ND solution has been automatically inserted and agitation begins, manually add 20g of Na2SO3 and 4.0 ml of alpha-amylase.
20g``4.0mL
Equipment
| Value | Label |
|---|---|
| Fiber analyzer | NAME |
| Ankom | BRAND |
| A2000 | SKU |
Close the Vessel Lid.
Equipment
| Value | Label |
|---|---|
| Fiber analyzer | NAME |
| Ankom | BRAND |
| A2000 | SKU |
This step takes about 1h 30m 0s.
When the NDF extraction and rinsing procedures are complete, open the Vessel Lid and remove the filter bags. Gently press out excess water from the bags.
Equipment
| Value | Label |
|---|---|
| Fiber analyzer | NAME |
| Ankom | BRAND |
| A2000 | SKU |
Completely dry the filter bags in an oven at 102 ± 2°C for 3 h.
3h 0m 0s
During that time, go to Step 20 and Flush Procedure (Step 21).
Equipment
| Value | Label |
|---|---|
| Oven | NAME |
| Oven forced-air convection | TYPE |
| Fisher Isotemp | BRAND |
| 15-103-0510 | SKU |
During that time, run a flush procedure if necessary.
Detach both cubitainers from ports A and B.
Flush procedure:
Verify that the drain hose is securely inserted into a waste bin.
Flush procedure:
Press the arrow keys on the Keypad until you see “Select Flush” on the Display.
Flush procedure:
Press ENTER on the Keypad.
Flush procedure:
Attach the Amylase Dispenser Assembly to port B.
Flush procedure:
Fill the dispenser with 100 ml of hot water.
Flush procedure:
Attach the Amylase Dispenser Assembly to port A.
Flush procedure:
Fill the dispenser with about 100 ml of hot water.
When dry, remove the filter bags from the oven and immediately place them directly into a collapsible desiccant pouch and flatten it to remove any air. Cool to ambient temperature for 0h 30m 0s.
Weigh the filter bags to the nearest 0.0001 g and add record the weight (in Fulcrum).
Remove the yellow waste bin and label it properly.
Acid Detergent Fiber
Place up to 3 bags on each of eight Bag Suspender Trays (maximum of 24 bags). Stack the trays on the center post of the Bag Suspender with each level rotated 120 degrees in relation to the tray below it. Place the empty 9th tray on top.
Insert the drain hose securely into an empty white waste bin.
Read the Temperature Controller on the right side of the instrument.
Attach the acid detergent (AD) solution hose to the Cubetainer and then to Port B on the instrument.
Open the Vessel Lid and insert the Bag Suspender with bags into the Vessel and place the Bag Suspender Weight on top of the empty 9th tray to keep the Bag Suspender submerged.
Follow the instructions on the ANKOM2000 display:
a. Select ADF.
b. Close the Vessel Lid.
c. Confirm hot water is on (>70°C).
d. Press START .
1h 0m 0s
When the ADF extraction and rinsing procedures are complete, open the Vessel Lid and remove the filter bags. Gently press out excess water from the bags.
Completely dry the filter bags in an oven at 102 ± 2°C for 3 h.
3h 0m 0s
Equipment
| Value | Label |
|---|---|
| Oven | NAME |
| Oven forced-air convection | TYPE |
| Fisher Isotemp | BRAND |
| 15-103-0510 | SKU |
During that time, remove the white waste bin and label it properly.
During that time, run a flush procedure if necessary.
Detach both cubitainers from ports A and B.
Flush procedure:
Verify that the drain hose is securely inserted into a waste bin.
Flush procedure:
Press the arrow keys on the Keypad until you see “Select Flush” on the Display.
Flush procedure:
Press ENTER on the Keypad.
Flush procedure:
Attach the Amylase Dispenser Assembly to port B.
Flush procedure:
Fill the dispenser with 100 ml of hot water.
Flush procedure:
Attach the Amylase Dispenser Assembly to port A.
Flush procedure:
Fill the dispenser with 100 ml of hot water.
Remove the filter bags from the oven, immediately place them into a collapsible desiccant pouch, and flatten it to remove any air. Cool to ambient temperature.
Weigh the filter bags to the nearest 0.0001 g and record the weight (in Fulcrum).
Equipment
| Value | Label |
|---|---|
| Sartorius Practum | NAME |
| Analytical balance | TYPE |
| Sartorius | BRAND |
| 224-1S | SKU |
Lignin method in beakers
Prepare a 72% H2SO4 solution or take a 250 ml bottle of 72% H2SO4.
.
Place dried filter bags with samples into a 2 L glass beaker and completely cover the bags with 72% H2SO4 (approximately 250 ml).
Place a 1 L beaker inside the 2 L beaker to keep the filter bags submerged.
Agitate the bags at the start and at 30-minute intervals by gently pushing and lifting the 1 L beaker up and down approximately 30 times.
After 3 h, pour off the H2SO4 and rinse with tap water to remove all of the acid.
Repeat rinses until the pH paper shows neutral color when touching the bags. Handle the bags gently during rinsing. Fine lignin particles can exit the filter bag if not handled carefully.
Dry the bags in an oven at 100 °C for 3 h.
3h 0m 0s
Equipment
| Value | Label |
|---|---|
| Oven | NAME |
| Oven forced-air convection | TYPE |
| Fisher Isotemp | BRAND |
| 15-103-0510 | SKU |
Remove the filter bags from the oven and place them directly into a Desiccant/Moisture Stop pouch. Flatten the pouch to remove air. Cool to ambient temperature and weigh the bags to the nearest 0.0001 g. Record the weight (in Fulcrum).
Ashing
Rinse 24 crucibles and covers with dH2O.
Dry the crucibles and covers in an oven at 150 °C for 1 h.
1h 0m 0s
Equipment
| Value | Label |
|---|---|
| Oven | NAME |
| Oven forced-air convection | TYPE |
| Fisher Isotemp | BRAND |
| 15-103-0510 | SKU |
Cool down crucibles (with lids on) in a metal dessicator for 1h30min.
1h 30m 0s
Equipment
| Value | Label |
|---|---|
| Fisherbrand™ Stainless Steel Desiccator with Stainless Steel Shelves | NAME |
| Metal Dessicator | TYPE |
| Fisher | BRAND |
| S02123 | SKU |
Weigh empty crucibles, with lids, to the nearest 0.0001 g. Record the weight (in Fulcrum).
Equipment
| Value | Label |
|---|---|
| Sartorius Practum | NAME |
| Analytical balance | TYPE |
| Sartorius | BRAND |
| 224-1S | SKU |
Fold filters bags into four and insert each bag into its crucible, with the corresponding sample number. Cover with the lid.
Put the crucibles (with lids on) in the muffle furnace at 500 °C for 5 h.
5h 0m 0s
Equipment
| Value | Label |
|---|---|
| Thermo Scientific™ Thermolyne™ Largest Tabletop Muffle Furnaces | NAME |
| Muffle furnace | TYPE |
| Thermolyne | BRAND |
| 10-505-13 | SKU |
Change the temperature at 150 °C, O/N.
14h 0m 0s
Equipment
| Value | Label |
|---|---|
| Thermo Scientific™ Thermolyne™ Largest Tabletop Muffle Furnaces | NAME |
| Muffle furnace | TYPE |
| Thermolyne | BRAND |
| 10-505-13 | SKU |
Turn off the muffle furnace and transfer the crucibles (with lids on) in a metal dessicator for 1h30min.
1h 30m 0s
Equipment
| Value | Label |
|---|---|
| Fisherbrand™ Stainless Steel Desiccator with Stainless Steel Shelves | NAME |
| Metal Dessicator | TYPE |
| Fisher | BRAND |
| S02123 | SKU |
Weigh crucibles with lids on to the nearest 0.0001 g. Record the weight (in Fulcrum).
Equipment
| Value | Label |
|---|---|
| Sartorius Practum | NAME |
| Analytical balance | TYPE |
| Sartorius | BRAND |
| 224-1S | SKU |
Wash crucibles and lids with tap water.