云顶集团3118(中国)

    MicroCT protocols for scanning of embryos and juvenile Hexaplex trunculus

    Eva Chatzinikolaou, Kleoniki Keklikoglou

    Published: 2021-09-03 DOI: 10.17504/protocols.io.bxw4ppgw
    AI 解读
    浏览 1310

    Disclaimer

    DISCLAIMER – FOR INFORMATIONAL PURPOSES ONLY; USE AT YOUR OWN RISK

    The protocol content here is for informational purposes only and does not constitute legal, medical, clinical, or safety advice, or otherwise; content added to protocols.io is not peer reviewed and may not have undergone a formal approval of any kind. Information presented in this protocol should not substitute for independent professional judgment, advice, diagnosis, or treatment. Any action you take or refrain from taking using or relying upon the information presented here is strictly at your own risk. You agree that neither the Company nor any of the authors, contributors, administrators, or anyone else associated with protocols.io, can be held responsible for your use of the information contained in or linked to this protocol or any of our Sites/Apps and Services.

    Abstract

    Micro-computed tomography (micro-CT) is a high-resolution 3D-imaging technique which is now increasingly applied in biological studies focusing on taxonomy and functional morphology. The creation of virtual representations of specimens can increase availability of otherwise underexploited and inaccessible samples. This protocol aims to standardise micro-CT scanning procedures for embryos and juveniles of the marine gastropod species Hexaplex trunculus.

    Steps

    Sample preparation

    1.

    Removal of embryos of Hexaplex trunculus from their egg capsules.

    2.

    Anesthetization of embryos and juvenile gastropods with 7% MgCl2.

    3.

    Sample placement inside a plastic white pipette tip without any scanning medium (in the air).

    microCT scanning

    4.

    Equipment

    ValueLabel
    SkyScan 1172NAME
    micro-computed tomographer (microCT)TYPE
    BrukerBRAND
    10L01170SKU
    Detail detectability: <1 μm
    Low contrast resolution (10% MTF): 5 μm
    Pixel size at maximum magnification: <0.8 μm

    X-ray source: Sealed microfocus X-ray tube, air cooled, >10,000h lifetime; Spot size <5μm @ 4W, 20-100kV, 0-250 μA (10W max)

    X-ray detector (camera): 11 Megapixel (4000X2300) 12-bit digital CCD-camera with fibre optic coupling to scintillator

    Maximum object size: 50 mm in diameter using offset scan

    Radiation safety: < 1 µSv/h at any point on the instrument surface | SPECIFICATIONS |

    4.1.

    Scanning parameters for embryos and juvenile Hexaplex trunculus

    Voltage: 59 kV

    Current :167 μA

    Filter: none

    Pixel size: 2 μm

    Camera binning: 1 × 1

    Exposure time: 325 ms for embryos and 316 ms for juveniles

    Rotation: 360° for embryos and 180° for juveniles

    Rotation step: 0.20º for embryos and 0.25° for juveniles

    Frame averaging: 3 for embryos and 5 for juveniles

    Images reconstruction

    5.

    Projection images were reconstructed into cross sections using the SkyScan’s NRecon software (Bruker, Kontich, Belgium)

    5.1.

    Upload projection images

    5.2.

    Perform X-Y alignment

    5.3.

    Reconstruction parameters

    Smoothing: 2

    Ring artifact correction: 20

    Beam hardening correction: 59%

    Attenuation coefficients: 0 - 1.127 for embryos and 0 - 1 for juveniles

    Save as: 16-bit TIFF images

    推荐阅读

    Nature Protocols
    Protocols IO
    Current Protocols
    扫码咨询