• 云顶集团3118(中国)

    Pilot Exercise: Generating the Illumina SampleSheet and sharing data via BaseSpace

    Ruth Timme, Maria Balkey, Tunc Kayikcioglu, Candace Bias, Cameron Boerner, James Pettengill

    Published: 2021-12-03 DOI: 10.17504/protocols.io.b2i3qcgn
    AI 解读
    浏览 1367

    Abstract

    This protocol describes the required steps to generate Illumina SampleSheet and provides guidelines for sharing data via BaseSpace with FDA for the 2021 Pilot Exercise

    V2:

    • added more details regarding data transfer through BaseSpace, including steps for sharing a project with FDA-CFSAN.

    • replaced samplesheet example.

    Steps

    Generating Sequencing SampleSheet

    1.

    Create your SampleSheet using Excel or a text editor. Name your sample sheet according your internal protocols, and use a *.csv extension. SampleSheet.csv The sample sheet is organized in sections titled Header, Reads, Settings and Data. Section headings are case-sensitive and shown in brackets [ ]. Details for each section are described below.

    1.1.

    Header Section.

    AB
    ParameterDescription
    Investigator NameFirst Name and Last Name for laboratory scientist
    Experiment NamePilot Exercise SARS-Cov-2
    DateYYYY-MM-DD
    WorkflowGenerateFASTQ
    ApplicationFASTQ Only
    Instrument TypeMiSeq
    Library Prep KitSwift Normalase Amplicon NGS Panels (SNAP)
    Index KitSwift Normalase Amplicon NGS Panels (SNAP)
    DescriptionFDA Wastewater Protocol Pilot Exercise
    ChemistryAmplicon

    Table 1: Header Section in Sequencing SampleSheet

    1.2.

    Reads Section

    **Enter the length of reads you are generating (could be different from 151)

    AB
    ParameterDescription
    Number of cylces in Read 1151
    Number of cycles in Read 2151

    Table 2: Reader Section in Sequencing SampleSheet

    1.3.

    Settings Section

    AB
    ReverseComplement0

    Table 3: Settings Section in Sequencing SampleSheet

    1.4.

    Data Section: Include the machine ID in the Sample_Name column. e.g. sample_B01_M01234 where M01234 corresponds to the machine ID. This is the ID that will get populated to the fastq file names.

    ABCDEFGHIJ
    Sample_IDSample_NameSample_PlateSample_WellI7_Index_IDindexI5_Index_IDindex2Sample_ProjectDescription
    WPP_sample_B.01WPP_sample_B.01_M012341A1SU001TTGTTCCTTGI5_1CCGAACAACALabAbreviation-WPP
    WPP_sample_C.01WPP_sample_C.01_WPP-sample_B.011B1SU002AGGTTGTGTTI5_2AAGAATCGGCLabAbreviation-WPP
    WPP_sample_SA-1.01WPP_sample_SA-1.01_WPP-sample_B.011C1SU003TATTGGTTGGI5_3CACTTCGCTTLabAbreviation-WPP
    WPP_sample_SA-2.01WPP_sample_SA-2.01_WPP-sample_B.011D1SU004CACTTCGCTTI5_4TATTGGTTGGLabAbreviation-WPP

    Table 4: Data Section in Sequencing SampleSheet (dashes converted to underscores for Illumina requirements)

    1.5.

    If a new library or sequencing protocol is utilized for processing the pilot exercise samples, make the changes in the corresponding parameters for the sequencing sample sheet.

    2.

    Starting your MiSeq run and connecting to BaseSpace:

    When you set up the run on the MiSeq, select the option to log in to BaseSpace.

    Transfer sequencing RUN through BaseSpace

    3.

    Click the Runs tab in the Illumina BaseSpace website .

    3.1.

    Select the run that you would like to share with the FDA LFFM wastewater project team.

    3.2.

    Go to the summary tab and click at the share button.

    3.3.

    Enter the email address for the FDA team ( covidtrakr@fda.hhs.gov) , and then click Add Collaborator.

    3.4.

    Click Save Settings. Your run will be automatically shared with the FDA LFFM wastewater project team.

    Figure 1: Sharing data via BaseSpace by Email Invitation
    Figure 1: Sharing data via BaseSpace by Email Invitation

    Transfer sequencing PROJECT through BaseSpace

    4.

    Click the Projects tab in the Illumina BaseSpace website .

    4.1.

    Select the project that you would like to share with the FDA LFFM wastewater project team.

    4.2.

    Click at the share project button.

    4.3.

    Enter the email address for the FDA team ( covidtrakr@fda.hhs.gov) , and then click Add Collaborator.

    4.4.

    Click Save Settings. Your project will be automatically shared with the FDA LFFM wastewater project team.

    推荐阅读

    Nature Protocols
    Protocols IO
    Current Protocols